Kwik

PA-2 Mercury Process Analyzer
Continuous Detection of Mercury in Water and Aqueous Solutions

 

Mercury Instruments

Mercury Instruments

Fully automatic system

Mercury Instruments

Easy menu-driven operation

Mercury Instruments

Proven and reliable detection method: UV absorption

Mercury Instruments

Measuring ranges from 0.01 ppb to 100 ppm

Mercury Instruments

High flexibility of sample pretreatment procedure

Mercury Instruments

Right for complex sample compositions

Mercury Instruments

Low reagent consumption

Mercury Instruments

Corrosion-protected construction

Mercury Instruments

Automatic self diagnosis system for reliable operation

 

 

Mercury Instruments

On-line Process Control of Mercury with the PA-2


The mercury process analyzer PA-2 is used for continuous monitoring of mercury concentrations in industrial processes. Applications include effluent and quality control in chlorine-alkali plants, monitoring of scrubber water of waste incinerators and power plants, control of industrial sewage and purification plants, quality control of sulphuric acid, drinking water. Special version also for concentrated caustic solutions (NaOH, KOH)

Online Measurement of mercury in concentrated (50%) caustic soda with the PA-2 (left), the sample passes a specially designed dilution system (right). Location: chlor-alkali plant (France).

Mercury Instruments

 

 

Mercury Instruments

Flexibility in Sample Digestion

In practice mercury analysis sample matrices very often show different and variable compositions. Depending on the chemical process the forms of mercury differ: elementary, ionic, dissolved as a salt, as an organic compound or as insoluble sulfide.

The PA-2 offers a variety of sample pretreatment procedures to allow highly precise determination of total mercury in all these samples, with the results from the PA-2 showing very good correlations with standard laboratory analyses.

High flexibility is achieved by the modular structure of the sample pretreatment module. This allows the user to choose from the following digestion methods:

http://www.mercury-instruments.de/EN/navigation/bullet-blue.gif

Sample preoxidation with hydrochloric acid + potassium permanganate

Mercury Instruments

Sample preoxidation with sulphuric acid + potassium permanganate

Mercury Instruments

Sample preoxidation with hydrochloric acid + sodiumchlorate

Mercury Instruments

Sample preoxidation with Fenton´s reagent

 

 

The integrated heated reaction module allows accelerated sample digestion at increased temperatures. The duration of reactions can be adapted to the need of your application by employing reaction coils of different lengths.

 

Mercury Instruments

Proven Measurement Principle

The detection of mercury contained in the sample occurs in an optical cell made of fused silica (Suprasil). In a first step mercury is reduced to the elementary state by means of tin(II)chloride or NaBH4. Subsequently mercury is stripped from the aqueous phase with an air stream and carried into the optical cell. Here the UV absorption measurement is conducted at a wavelength of 253,7 nm.

This analytical technique called "cold vapor atomic absorption spectroscopy (CVAAS)" shows extreme sensitivity and selectivity. It has been reliable and proven for many years. In contrast to the occasionally recommended atomic fluorescence method the analytical technique utilized by the PA-2 is extremely low in interference and does not require an amalgamation step nor expensive noble gas as a carrier.

 

Mercury Instruments

Industrial Grade Design

To provide optimum protection against corrosive environments all parts of the PA-2 Mercury Analyzer are enclosed in an industrial-grade cabinet made of fibreglass-reinforced polyester (protection class IP 66; NEMA 4X; etc.). The electronic circuitry is shielded from the wet chemical section by a chemically resistant wall.

http://www.mercury-instruments.de/EN/products/images/PA-1_Application.JPG

http://www.mercury-instruments.de/EN/products/images/PA-1_ECI.jpg

PA-2 for continuous monitoring of
Hg in scrubber water (coal fired
power station, Germany)

PA-2 reading the Hg concentration of
effluents (chlor-alkali plant, Germany)

 

Continuous and Interval Operation

The PA-2 analyzer is controlled by a built-in computer and operates fully automatic. The continuous measuring mode can be switched over to a periodic measuring mode (for example 10 minutes of measurement every hour). Auto-Cal. Check: periodically the instrument switches from the sample stream to a calibration solution to check any deviation from the reference value. A status signal is generated if the value measured is out of range. Auto-Zero: the zero line of the analyzer is adjusted automatically after a preset period of time. Self-Cleaning: Precipitations like manganese dioxide are dissolved by an automatic rinse step which makes manual cleaning superfluous.

 

Easy to Operate


The PA-2 is operated via a waterproof membrane keypad. All inputs required are selected in a readily understandable menu shown on the graphical display. It is also possible to initiate some functions like Auto Zero from an external computer via the RS 232 interface.

 

Display and Output of Measurements

The analytical results are continuously displayed on an LCD in micrograms per litre and as absorption units. A linear electrical signal (4 ... 20 mA) is also output and can be connected to a recorder or to the analog input of a data aquisition system.

 

Automatic Self Diagnosis System

The PA-2 is equipped with sensors to detect malfunctions of the system and triggers an alarm for the operator. The following functions/malfunctions are checked and indicated:

http://www.mercury-instruments.de/EN/navigation/bullet-blue.gif

Reagent flow

Mercury Instruments

Sample flow

Mercury Instruments

Stripping air flow

Mercury Instruments

Leakage of fluids

Mercury Instruments

Photometer lamp

Mercury Instruments

Calibration

 

 

These features permit reliable unattended operation. Malfunctions are reported to the user both optically and via a status signal.

 

 

Minimum Maintenance Work and Maximum Service Life

The PA-2 is not a laboratory analyzer simply converted for process applications but has especially been designed for operation under harsh industrial conditions. The number of parts subject to wear has been minimized and particularly durable components have been chosen for use in critical locations. This results in extended maintenance intervals. The stripping unit is based on an aerosol-free principle and the need for cleaning the optical cell is particularly low.

 

Communication with External Computers

The following data is available at the serial RS 232 interface: Actual mercury concentration Status (Ready - Maintenance - Malfunction)

Thanks to the bidirectional interface, operational functions like "Zero Adjustment", "Calibration Check" and "Start Measurement" can be triggered externally

 

Mercury Instruments

PA-2 Mercury Process Analyzer Technical Specifications

Measuring principle:

UV-Absorption

Wavelength:

254 nm

UV source:

Electrodeless low-pressure mercury lamp

Stabilization method:

Double beam (reference beam) method

Optical cell:

Fused silica (Suprasil)

Optical cell heating:

approx. 70°C

Measuring range:

lowest range: 0...1 µg/L
highest range: 0...100 mg/l

Response time (99% signal):

approx. 1 minute

Carrier gas:

Air, approx. 30 L/h, 1 bar

Pump for liquids:

Multichannel peristaltic pump

Reduction reagent:

Tin-II-chloride or alternatively sodiumhydroborate

Reagent consumption:

approx. 3 l / week for each reagent

Sample digestion:

HCl or H2SO4, KMnO4 or H2O2 or Fenton´s Reagent or Na-chlorate (depending on sample composition)

Liquid-gas separation:

Aerosol-free principle

Input device:

Waterproof membrane keypad

Display of readings:

Graphical LC display with background illumination

Output for concentration:

4...20 mA; RS 232 bidirectional

Output for status:

Operational - Service - Malfunction, according to NAMUR 64 recommendations

Housing:

Fibreglass-reinforced polyester, for use in highly corrosive environment - Reagent storage cabinet made of chemically resistent plastic material

Protection class:

IP 66 (EN 60529/ NEMA 3; 3R; 4; 4X; 12; 13)

Power Supply:

230 V / 50 Hz (115 V / 60 Hz option)

Power consumption:

250 W max.

Air consumption:

Instrument air, approx. 30 l / h,
1 - 6 bar (14 - 85 psi)

Dimensions:

775 x 612 x 329 mm (WxHxD)

Weight:

approx. 50 kg

 

 

 